Comparative analysis of anterior corneal curvature and astigmatism measurements obtained with three different devices

Empreu sempre aquest identificador per citar o enllaçar aquest ítem http://hdl.handle.net/10045/109393
Información del item - Informació de l'item - Item information
Títol: Comparative analysis of anterior corneal curvature and astigmatism measurements obtained with three different devices
Autors: Molina-Martín, Ainhoa | Piñero, David P. | Caballero, María T. | Fez Saiz, Dolores de | Camps, Vicente J.
Grups d'investigació o GITE: Grupo de Óptica y Percepción Visual (GOPV)
Centre, Departament o Servei: Universidad de Alicante. Departamento de Óptica, Farmacología y Anatomía
Paraules clau: Astigmatism | Cassini | Colour light-emitting diode reflection topography | Cornea | Corneal topography | IOL master | Keratometry | Pentacam
Àrees de coneixement: Óptica
Data de publicació: de setembre-2020
Editor: Wiley
Citació bibliogràfica: Clinical and Experimental Optometry. 2020, 103(5): 618-624. https://doi.org/10.1111/cxo.13002
Resum: Background: The objective of this study was to compare the central corneal curvature and astigmatism measurements obtained with three different systems in healthy eyes and to assess the level of interchangeability between them. Methods: This was a comparative study examining 30 healthy eyes of 30 patients (age 15–53 years). A complete eye examination was performed in all cases including analysis of anterior corneal curvature and astigmatism with three devices: the colour‐LED topography system Cassini (i‐Optics) (CAS), the Scheimpflug‐based system Pentacam (Oculus Optikgeräte) (PTC) and the optical biometer IOL‐Master 500 (Carl Zeiss Meditec) (IOLM). Differences between devices in terms of curvature in the flattest (flat K) and steepest meridians (steep K) as well as in the magnitude (AST) and power vector components of astigmatism (J0 and J45) were evaluated. The interchangeability between devices was evaluated with the Bland–Altman method. Results: Statistically significant differences between devices were found in steep K and flat K (p < 0.001). No statistically significant differences between devices were found in AST (p = 0.057) and J0 power vector (p = 0.185). However, differences between devices in J45 did reach statistical significance (p = 0.039). Ranges of agreement for curvature measures ranged from 0.123 (flat K, CAS‐PTC) to 0.165 mm (steep K, CAS‐PTC). Ranges for the magnitude of astigmatism were 0.868, 1.059 and 0.739 D in CAS‐IOLM, PTC‐IOLM and CAS‐PTC comparisons, respectively. For J0 and J45, ranges of agreement were below 0.522 D. Conclusions: Measurements of central corneal curvature and astigmatism obtained with the three devices evaluated cannot be used interchangeably.
Patrocinadors: David P. Piñero has been supported by the Ministry of Economy, Industry and Competitiveness of Spain within the program Ramón y Cajal, RYC-2016-20471.
URI: http://hdl.handle.net/10045/109393
ISSN: 0816-4622 (Print) | 1444-0938 (Online)
DOI: 10.1111/cxo.13002
Idioma: eng
Tipus: info:eu-repo/semantics/article
Drets: © 2019 Optometry Australia
Revisió científica: si
Versió de l'editor: https://doi.org/10.1111/cxo.13002
Apareix a la col·lecció: INV - GOPV - Artículos de Revistas

Arxius per aquest ítem:
Arxius per aquest ítem:
Arxiu Descripció Tamany Format  
ThumbnailMolina-Martin_etal_2020_ClinExpOptom_final.pdfVersión final (acceso restringido)263,56 kBAdobe PDFObrir     Sol·licitar una còpia


Tots els documents dipositats a RUA estan protegits per drets d'autors. Alguns drets reservats.