Confounding sizing in posterior chamber phakic lens selection due to white-to-white measurement bias
Por favor, use este identificador para citar o enlazar este ítem:
http://hdl.handle.net/10045/88412
Título: | Confounding sizing in posterior chamber phakic lens selection due to white-to-white measurement bias |
---|---|
Autor/es: | Fernández, Joaquín | Rodríguez-Vallejo, Manuel | Martínez, Javier | Tauste Francés, Ana | Hueso, Elisa | Piñero, David P. |
Grupo/s de investigación o GITE: | Grupo de Óptica y Percepción Visual (GOPV) |
Centro, Departamento o Servicio: | Universidad de Alicante. Departamento de Óptica, Farmacología y Anatomía |
Palabras clave: | Agreement | Posterior chamber phakic intraocular lens | Reproducibility | Sizing | White-to-white |
Área/s de conocimiento: | Óptica |
Fecha de publicación: | 18-feb-2019 |
Editor: | Wolters Kluwer - Medknow |
Cita bibliográfica: | Indian Journal of Ophthalmology. 2019, 67(3): 344-349. doi:10.4103/ijo.IJO_613_18 |
Resumen: | Purpose: To assess the agreement in the white-to-white (WTW) measurement with two different devices, the reproducibility and the probability of confusing sizing (PCS) in selecting a different implantable collamer lens (ICL). Study Design: Retrospective observational case series. Methods: Images of 192 eyes were captured with both devices. The WTW was measured automatically (OA) and manually (OM) with the Orbscan and Keratograph (KA and KM) by one examiner who repeated a total of four measures. A second examiner conducted a single manual measure for each device over the same image. The ICL sizing was computed for each measure of WTW and the PCS was calculated as the percentage of cases for which the confronted or repeated measure resulted in a different size of the ICL. The critical WTWs with highest PCS were identified. Results: KM overestimated the WTW versus OM in 0.13 ± 0.18 mm (P < 0.001) but not in the automated method comparison, 0.01 ± 0.19 mm (P = 0.58). Inter-examiner reproducibility (R) was higher with OM than with KM, and the intra-examiner R decreased with the average of two measures in both cases. The PCS was higher with the increase of mean differences, the limits of agreement (LoAs), and R. WTWs from 11.1 to 11.2 mm, 11.6 to 11.7 mm, and 12.3 to 12.4 mm resulted in higher PCS. Conclusion: The mean difference is not enough to apply conversions between devices and the LoAs and R should be considered. Special attention should be taken for WTWs with higher PCS. |
URI: | http://hdl.handle.net/10045/88412 |
ISSN: | 0301-4738 (Print) | 1998-3689 (Online) |
DOI: | 10.4103/ijo.IJO_613_18 |
Idioma: | eng |
Tipo: | info:eu-repo/semantics/article |
Derechos: | © 2019 Indian Journal of Ophthalmology | Published by Wolters Kluwer - Medknow. This is an open access article distributed under the terms of the Creative Commons Attribution‑NonCommercial‑ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non‑commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
Revisión científica: | si |
Versión del editor: | https://doi.org/10.4103/ijo.IJO_613_18 |
Aparece en las colecciones: | INV - GOPV - Artículos de Revistas |