Biometric Factors Associated with the Visual Performance of a High Addition Multifocal Intraocular Lens

Por favor, use este identificador para citar o enlazar este ítem: http://hdl.handle.net/10045/77795
Información del item - Informació de l'item - Item information
Título: Biometric Factors Associated with the Visual Performance of a High Addition Multifocal Intraocular Lens
Autor/es: Fernández, Joaquín | Rodríguez-Vallejo, Manuel | Martínez, Javier | Tauste Francés, Ana | Piñero, David P.
Grupo/s de investigación o GITE: Grupo de Óptica y Percepción Visual (GOPV)
Centro, Departamento o Servicio: Universidad de Alicante. Departamento de Óptica, Farmacología y Anatomía
Palabras clave: Multifocal intraocular lens | Visual acuity defocus curves | Corneal power | Lens centration | Effective addition
Área/s de conocimiento: Óptica
Fecha de publicación: 30-may-2018
Editor: Taylor & Francis
Cita bibliográfica: Current Eye Research. 2018, 43(8): 998-1005. doi:10.1080/02713683.2018.1478981
Resumen: Purpose/Aim: To evaluate the impact of ocular parameters on the visual performance achieved with the multifocal intraocular lens (IOL) Bi-Flex M 677MY. Materials and Methods: About 26 subjects were included in the current study. Several physiological variables were retrieved from the 3-month follow-up visit, including pupil diameter and distance from pupil center to the vertex normal of the anterior cornea (µ). These variables were also obtained in the preoperative visit. Binocular and monocular visual acuity defocus curves were measured at 1 and 3 months after surgery, respectively. The monocular Area Under the Curve (AUC) was computed along the total range (Total Area Under the Curve (TAUC), +1.00 to −4.00 D) and for the ranges of Far (Far Area Under the Curve (FAUC), +0.50 to −0.50 D), (Intermediate Area Under the Curve (IAUC), −1.00 to −1.50 D) and Near vision (Near Area Under the Curve (NAUC), −2.00 to −4.00 D). Correlations between these areas and the postoperative physiological variables were assessed. Results: The mean µ was reduced from 0.21 to 0.10 mm after surgery, as well as pupil diameters, either photopic (−7.4%) or mesopic (−8.1%) (p < 0.05). The mean AUCs were 2.08 ± 0.74 for TAUC, 0.57 ± 0.17 for FAUC, 0.16 ± 0.09 for IAUC, and 0.81 ± 0.29 for NAUC. Significant correlations were found between NAUC and corneal power (r = −0.39, p = 0.05) as well as between IAUC and temporal decentration of the lens from vertex normal (ρ = −0.41, p = 0.04). Conclusions: The visual performance at near distance with the IOL evaluated improved in eyes with less corneal power. On the other hand, a slight temporal IOL decentration from vertex normal also improved intermediate visual acuity. The binocular defocus curve was similar to other trifocal IOLs.
URI: http://hdl.handle.net/10045/77795
ISSN: 0271-3683 (Print) | 1460-2202 (Online)
DOI: 10.1080/02713683.2018.1478981
Idioma: eng
Tipo: info:eu-repo/semantics/article
Derechos: © 2018 Taylor & Francis Group, LLC
Revisión científica: si
Versión del editor: https://doi.org/10.1080/02713683.2018.1478981
Aparece en las colecciones:INV - GOPV - Artículos de Revistas

Archivos en este ítem:
Archivos en este ítem:
Archivo Descripción TamañoFormato 
Thumbnail2018_Fernandez_etal_CurrentEyeRes_final.pdfVersión final (acceso restringido)1,22 MBAdobe PDFAbrir    Solicitar una copia
Thumbnail2018_Fernandez_etal_CurrentEyeRes_accepted.pdfAccepted Manuscript (acceso abierto)1,47 MBAdobe PDFAbrir Vista previa


Todos los documentos en RUA están protegidos por derechos de autor. Algunos derechos reservados.