Simple-to-use and portable device for free chlorine determination based on microwave-assisted synthesized carbon dots and smartphone images

Empreu sempre aquest identificador per citar o enllaçar aquest ítem http://hdl.handle.net/10045/113773
Información del item - Informació de l'item - Item information
Títol: Simple-to-use and portable device for free chlorine determination based on microwave-assisted synthesized carbon dots and smartphone images
Autors: Uriarte, Damián | Vidal, Ezequiel | Canals, Antonio | Domini, Claudia E. | Garrido, Mariano
Grups d'investigació o GITE: Espectroscopía Atómica-Masas y Química Analítica en Condiciones Extremas
Centre, Departament o Servei: Universidad de Alicante. Departamento de Química Analítica, Nutrición y Bromatología | Universidad de Alicante. Instituto Universitario de Materiales
Paraules clau: Portable device | Smartphone | Carbon dots | Fluorimetric determination | Free chlorine
Àrees de coneixement: Química Analítica
Data de publicació: 1-de juliol-2021
Editor: Elsevier
Citació bibliogràfica: Talanta. 2021, 229: 122298. https://doi.org/10.1016/j.talanta.2021.122298
Resum: A new portable and simple 3D printed device was designed for free chlorine determination in water samples. The analytical method was based on the quenching caused by free chlorine on the fluorescence emission of the carbon dots (CD) synthesized from citric acid and urea. The fluorescence was captured through the camera of a smartphone, which was coupled to the 3D printed device, and the images were processed using the RGB system by the ImageJ 1.51q software. The proposed method was selective and precise (RSD% 4.6, for n = 6), and the trueness of the results was evaluated by comparing the results obtained with those recovered by the spectrophotometric method 4500-Cl G (standard method), with good agreement between them. Moreover, the remarkable correlation between the CD signal and the free chlorine concentration resulted in a determination with low detection limits (limit of detection of 6 μg L−1 and limit of quantification of 20 μg L−1). Therefore, the new method and the related portable device could be considered a fast, economical and reliable alternative for the on-site determination of free chlorine in water samples.
Patrocinadors: The authors wish to thank the financial support received from Universidad Nacional del Sur (PGI 24/Q099 and 24/Q086) and the ANPCyT (PICT-201-0659). The authors also thank Consejo Nacional de Investigaciones Científicas y Técnicas (CONICET) particularly for the doctoral grant of D. Uriarte.
URI: http://hdl.handle.net/10045/113773
ISSN: 0039-9140 (Print) | 1873-3573 (Online)
DOI: 10.1016/j.talanta.2021.122298
Idioma: eng
Tipus: info:eu-repo/semantics/article
Drets: © 2021 Elsevier B.V.
Revisió científica: si
Versió de l'editor: https://doi.org/10.1016/j.talanta.2021.122298
Apareix a la col·lecció: INV - SP-BG - Artículos de Revistas